Thermosel Temperature Control System

The Thermosel temperature control system allows precise viscosity measurements with fast temperature regulation using coaxial cylinders. Unlike a bath system, the Thermosel system does not use liquid for temperature control (with a maximum temperature of 300°C). It is compatible with standard Brookfield AMETEK rotational viscometers and the DVNEXT rheometer.

Letölthető brosúra

PDF

 
Kérdése van? Hívjon minket:
Máte záujem? Kontaktujte nás na čísle:
Have a question? Call us:

+36 30/228-3059

The Thermosel system, using coaxial cylinders, enables fast temperature regulation for precise viscosity measurements. Unlike a traditional bath system, the Thermosel does not use liquid for temperature regulation (with a maximum temperature of 300°C). It provides a stable and controlled environment for accurate viscosity measurement, essential for high-temperature materials such as hot melts and liquids.

By combining it with Brookfield viscometers, the Thermosel system enables precise and repeatable measurements. Key factors that contribute to a stable environment include:

  • Non-fluctuating temperature control.
  • Small sample volume and insulated chamber reduce temperature gradients within the sample.
  • Rotating spindle also acts as a mixer, ensuring even distribution of the sample.
  • Simple testing procedure that, once learned, allows for quick and reproducible data collection.

The system is compatible with standard Brookfield rotational viscometers and DVNXT rheometers. Please note: an optional DVP-141 cable is required.

With the Thermosel, temperature regulation is ensured up to +300°C. The system is also available with a magnetic spindle connection option.

Ramp execution between set values is possible using the RheocalcT (DVNXT & DV2T) software. An optional HT-106 cable is required for the Heat Tank (Heating Chamber).

Programmable temperature control is available for one or even up to ten points (at the desired temperature).

Direct temperature control is possible with the DV2T rotational viscometer or DVNXT rotational rheometer.


Thermosel Adapter Viscosity Ranges (cP, mPa·s):

Spindle Sample Volume Shear Rate (s-1) LVDVNXT LVDV2T LVDVE LVT RVDVNXT RVDV2T RVDVE RVT HADVNXT HADV2T HADVE HAT HBDVNXT HBDV2T HBDVE HBT
SC4-18 8 ml 1.32N 1.2-30K 1.5-30K 3-10K 5-10K 20-500K 25-500K 50-170K 50-100K 40-1M 50-1M 100-300K 100-200K 160-4M 200-4M 400-1.3M 400-800K
SC4-31 10 ml 0.34N 12-300K 15-300K 30-100K 50-100K 100-2.5M 125-2.5M 250-830K 250-500K 200-5M 250-5M 500-1.7M 500-1M 800-20M 1K-20M 2K-6.7M 2K-4M
SC4-34 9.5 ml 0.28N 24-600K 30-600K 60-200K 100-200K 200-5M 250-5M 500-1M 500-1M 400-10M 500-10M 1000-3.3M 1000-2M 1K-25M 2K-40M 4K-13.3M 4K-8M
SC4-34 8 ml 0.93N N/A N/A N/A N/A 500-12.5M 625-12.5M 1K-3.3M 1K-3.3M 1000-10M 1250-10M 2000-6.7M 2000-4M 2.5K-8.3M 3K-13.3M 6K-26.7M 8K-16M
SC4-27* 10.5 ml 0.34N 1.2-30K 1.5-30K 3-10K 5-10K 20-500K 25-500K 50-170K 50-100K 40-1M 50-1M 100-300K 100-200K 160-4M 200-4M 400-1.3M 400-800K
  • M = 1 million
  • K = 1 thousand
  • N = RPM
  • cP = Centi Poise
  • mPa·s = milli Pascal seconds
  • Note: Optional disposable SC4-27D spindle available in packs of 100 pieces, Product ID: SC4-27D-100. Special accessory required for connecting the disposable spindle to the viscometer.